  • For the RFID Technology to function, the syringe must be used with an Optivantage Injector with RFID technology. Function of the RFID technology is not dependent on syringe orientation as it is placed in the injector. Instructions for use of injector are provided on the injector interface screen and operator's manual.

    If the RFID tag is damaged or otherwise non-functional, the injector will notify the user. Should this occur the OPTIRAY syringe with the non-functional RFID tag may still be used but no data will be transferred to the injector.

    Regarding interference with medical devices, the RFID tag and injector system meet the IEC 60601-1-2 requirements for emission and immunity standards for medical devices. Follow all manufacturers' guidelines and do not operate any part of the Optivantage Injector System and RFID-tagged syringes within 6 inches (15 cm) of a pacemaker and/or defibrillator.

  • Storage: Store OPTIRAY and OPTIRAY RFID-tagged syringes at 25°C (77°F); excursions permitted to 15 to 30°C (59 to 86°F) [see USP Controlled Room Temperature]. OPTIRAY is sensitive to light and must be protected from strong daylight or direct exposure to the sun. If OPTIRAY syringes are frozen or if crystallization occurs, the syringe and contents should be discarded. If OPTIRAY in glass bottles is frozen or if crystallization occurs, the bottle and contents should be discarded. OPTIRAY may be stored at 40°C for up to one month in a contrast media warmer utilizing circulating warm air. When storing OPTIRAY for periods longer than one month, store at 25°C (77°F); excursions permitted to 15 to 30°C (59 to 86°F) [see USP Controlled Room Temperature]. Do not reautoclave plastic container because of possible damage to syringe.

    As with all contrast media, glass and plastic containers should be inspected prior to use to ensure that breakage or other damage has not occurred during shipping and handling. All containers should be inspected for closure integrity. Damaged containers should not be used.

    Optiray, Ultraject and Optivantage are trademarks of Mallinckrodt Inc.

    Mallinckrodt Inc.Hazelwood, MO 63042 USA

    MKR 13230308 Revised 0308 Printed in U.S.A.